Slit Lamp Microscope USER MANUAL
29 / 29
5.Trouble shooting
In case there is any trouble, please check according to the following table for reference. If it still
cannot work, please contact the after-sales service department of an authorized distributor.
Version: 1.0 20200610
Shanghai MediWorks Precision Instruments Co., Ltd.
No.7, MingPu Phase II, No.3279 SanLu Road,
MinHang District, Shanghai, China
Tel: 0086-21-54260421; Fax: 0086-21-54260425
Email: international@mediworks.biz
www.mediworks.com.cn
Illumination does not
light
The power cord is not properly
connected to the power outlet.
Connect the power cord correctly.
The main power switch is not turned
on.
Short press the switch key, the
power indicator light is on.
Loose plug on the power adapter
Tight plug on the power adapter
The lamp cover is not aligned
Replace the power adapter
The brightness adjustment knob is in
the Min
Adjustment brightness
adjustment knob
The reflect mirror coating is oxidized
Replace the reflect mirror
Excessive dust on the mirror surface
Clean the surface with soft
cotton
The brightness adjustment knob
position is not adjusted to the
maximum.
Adjust the brightness
adjustment knob.
The filter rod is located in the first or
middle position of the dimmer.
Correctly set the filter selector
lever position.
The slit is automatically
closed.
The slit width control button is too
loose.
Adjust the tightness of the
knob.
The fixation lamp does
not work..
The fixation lamp plug is loose.