-31-
Phenomenon Basic EMC
standard
Immunity test levels
Home Healthcare Environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field IEC 61000-4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3 Refer to table 3
Rated power frequency
magnetic fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
Table 2 - Enclosure Port