EasyManua.ls Logo

Megger KMK 8 - Page 72

Megger KMK 8
145 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-8
Having the measurement started the following display appears after a few
seconds and the bridge is ready for balancing.
Bridge Balancing
Select the minimum gain with key F2
Switch the measuring voltage off with the Vm OFF (F1) key and
keep it pressed until the pointer reaches a stable position. While
the key is pressed KMK 8 measures the disturbing DC voltage.
After releasing the key, KMK 8 performs a null-correction
compensating the effect of disturbing voltage. The null point
remains in the middle. Set the pointer to the 0 point with the M
balancing potentiometer
Increase the gain gradually with key F3 and repeat the first two
points until reaching the perfect balance. The balance is perfect
when the gain is 4 or 5 and the pointer stands on the 0 point with
pressed and released F1 key as well.
The device performs numerous measurements per second. The
pointer shows the result of the last measurement. When the line is
too noisy the pointer is waving around a middle point and the
reading is difficult. The waving of pointer can be reduced by
average calculation.
Five averaging times are provided: 0, 0.5, 1, 2 or 4 sec. (0= no
averaging)
The averaging time can be changed with keys F4 and F5. When
the balancing is completed press ENTER and the results appear.

Table of Contents

Related product manuals