11
Humidity
90% RH non-condensing at 40°C
Dimensions
305 x 194 x 360 (mm) (12 x 7.6 x 14.2 inches)
Weight
6.75 kg (15lb) approx
Voltage output accuracy (0°C to 30°C)
+4%, -0% ±10 V of nominal test voltage at 1 G load
Current measurement range
0.01 nA to 5 mA
Current measurement accuracy (23°C)
±5% ±0.2 nA at all voltages
Interference rejection
1 mA r
ms per 250 V test voltage up to a maximum of 2 mA
Timer range
Counts up to 99 minutes from star
t of test
Test regimes
Auto IR
Lead set
Three flexible silicon insulated leads with compact clamp.
Safety
Meets the requirements of EN61010-1:2001 CAT IV 600 V
EMC
Meets the requirements of EN61326-1:1998 for use in heavy industrial
areas.
Operating temperature
-20°C to 50°C
Storage temperature
-25°C to 65°C
Ingress protection (lid closed)
IP65