AF008-02-1201
82
Number
Issue
Date
CN Number
6
1
7
2-
999
2
17-01-07
22933
Compiled
By:
S.Mynott
Date
Compiled:
08/08/06
MIT5XX/10XX
1
st
Used On:
SERVICE MANUAL
Appendix 2 Firmware Error Codes.
Error Code Error Cause and description
1 to 11 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
12 Buzzer drive is unstable.
13 Buzzer drive pull down has failed.
14 EEPROM receive has failed.
15 EEPROM send has failed.
16 EEPROM did not acknowledge the command.
17 Non specific EEPROM error.
18 to 28 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
29-76 Not valid error codes
77 Incompatible display and measurement software
78 No communications with high voltage module.
79 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
80 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
81 Transmission to high voltage module failure
82 Measurement board comms reception time out.
83 to 91 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
92 Unknown type of high voltage board detected
93 No high voltage board detected
94 to 127 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
128 Test start key pressed during startup
129 to 146 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
147 Battery calibration result out of range
148 Battery calibration took too long
149 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
150 Capacitance measurement time out
151 to 156 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
157 Invalid model type
158 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
159 ADC running too fast
160 ADC running too slow
161 Unknown current filter on measurement PCB
162 to 165 Internal Software Error. Please report this to Megger with details on how to repeat the fault.
166 ADC not responding
167 to 181 Internal Software Error. Please report this to Megger with details on how to repeat the fault.

Other manuals for Megger MIT510