Maximum Current Output a.c.
PCITS2000/2 2000 A at 0 to 3 V, 50 Hz/60 Hz
1000 A at 0 to 6 V, 50 Hz/60 Hz
PCITS600/2 600 A at 0 to 4 V, 50 Hz/60 Hz
300 A at 0 to 8 V, 50 Hz/60 Hz
Auxiliary Voltage Output a.c.
0 to 250 V, 0-2 A, 50 Hz/60 Hz
0 to 125 V, 0-2 A, 50 Hz/60 Hz
All outputs are able to rise to 50 V above earth potential.
Timer Stop Inputs
Inputs for volt-free or a.c./d.c. energised contacts at
100 V to 264 V
Display
Dot matrix l.c.d.
Current Range PClTS 2000/2 0-2000 A, range resolution 1 A
Current Range PCITS 600/2 0-600 A, range resolution 1 A
Auxiliary Voltage Range 0-250 V, resolution 1 V; 0-2 A, resolution 0,01 A
Accuracy ±3% of reading ±2% LS digits
Elapsed Time 0-600 s, resolution 0,01 s, accuracy 0,1% ±0,05s
Duty Cycle
2
1
⁄2 minutes on circuit and 15 minutes off circuit at full
range current and voltage. Duty cycle increases until
continuous use is possible at 40% of full range.
Temperature Range
Operation 0˚C to 40˚C
Storage -20˚C to + 60˚C
Humidity Rang e
Operation (900% RH at 40˚C for 48 hours)
Storage (93% RH at 40˚C for 12 hours)
Supply Voltage
220 V/240 V +6%-10%, 50 Hz/60 Hz, single phase i.e.
198 V to 254 V at full output current.
7 kVA for PCITS2000/2, 3,5 kVA for PCITS600/2
Equipment Protection
Circuit Breakers 2 x 30 A rating breakers for PCITS2000/2, one short
delay and one long delay.
2 x 15 A rating breakers for PCITS600/2, one short delay
and one long delay.
Auxiliary Output Fuse 2 A ceramic HBC 20 mm x 5 mm IEC 127/1
Controller Supply Fuse 1 A ceramic HBC 20 mm x 5 mm IEC 127/1
Safety
The test set will, in general, meet the requirements of the
IEC 61010-1 specification
EMC
IEC 61326-1
Operational uncer tainties
Refer to www.megger.com
Dimensions
PCITS2000/2 320 mm x 305 mm x 510 mm (12
1
⁄2 in x 12 in x 20 in
approx.) excluding handles and wheels.
PCITS600/2 495 mm x 307 mm x 310 mm (19
1
⁄2) in x 12 in x 12
1
⁄4 in
approx.) excluding handles.
Controller 202 mm x 127 mm x 55 mm (8 in x 5 in x 2
1
⁄8 in approx.)
Weight
PCITS2000/2 61 kg (134 lb approx.)
PCITS600/2 38 kg (83
1
⁄2 lb approx.)
Cleaning
Wipe disconnected, cooled Test Set with a clean cloth
damped with soapy water or Isopropyl Alcohol (I.P.A.)
Specification