MD9272 Leakage Clamp TRMS Meter with Power Function Technical Specifications
20
5. Technical Specifications
General specifications · Environment conditions:
Measurement Category clamp-
meter
CAT III 600V
Measurement Category Test
Lead
CAT III 600V
Maximum Current
1
100 A
Pollution degree:
2
Altitude
< 2000 m
Operation temperature
0 ºC – 40 ºC (<80 % H.R., <10 ºC without
condensation)
Storage temperature
10 ºC – 60 ºC (<70 % H.R., extract batteries)
Temperature coefficient:
0.1 (of specified precision) / ºC (<18 º C or >28 º
C)
Functioning principle:
Sampling of signal at high speed
Data update on screen:
twice a second
Screen:
3 3/4 + 3 digits LCD screen
Range selection:
Automatic and manual selection
Over-range indication:
-0L- is shown on the LCD screen, fast beep
Polarity indication:
-“ is shown automatically
Power supply:
3 V
Type of batteries:
2 x 1,5 V AA, alkaline batteries
Consumption:
5 mA average, depending on function.
Duration of the batteries:
200 - 400 h. (batteries 2000 mAh)
Dimensions:
185 mm x 62 mm x 42 mm
Weight:
Aprox. 210 g (without battery)
(1) This equipment cannot be exposed to currents above the limit