3 Installation
846 Dosing Interface
■■■■■■■■
11
Figure 4 846 Dosing Interfaces with bottle holder 6.2061.010 or with
Titrando, respectively
3.3 Connecting the instrument to the power grid
WARNING
Electric shock from electrical potential
Risk of injury by touching live components or through moisture on live
parts.
Never open the housing of the instrument while the power cord is
still connected.
Protect live parts (e.g. power supply unit, power cord, connection
sockets) against moisture.
Unplug the power plug immediately if you suspect that moisture has
gotten inside the instrument.
Only personnel who have been issued Metrohm qualifications may
perform service and repair work on electrical and electronic parts.
Connecting the power cord
Accessories
Power cord with the following specifications:
Length: max. 2 m
Number of cores: 3, with protective conductor
Instrument plug: IEC 60320 type C13
Conductor cross-section 3x min. 0.75 mm
2
/ 18 AWG
Power plug:
according to customer requirement (6.2122.XX0)
min. 10 A