6.1 Mains connection
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
846 Dosing Interface
6 Technical specifications
6.1 Mains connection
Voltage 100…240 V (±10%)
Frequency 50…60 Hz
Power consump-
tion
maximum 45 W
Fuse Electronic overload protection
6.2 Interfaces and connectors
Controller connector
Controller port
USB upstream port with auxiliary power supply (Mini DIN socket) for
connecting Touch Control or computer for controlling the 846 Dosing
Interface.
Touch Control With integrated Touch Control cable
Computer with 6.2151.000 cable
USB connectors
USB ports
2 USB downstream ports (type A sockets), each 500 mA, for connect-
ing peripheral devices such as printers, keyboards, barcode readers or
RS-232/USB boxes (Metrohm order no. 6.2148.020).
Controller port USB upstream port with auxiliary power supply (Mini DIN socket) for
connecting Touch Control or computer for controlling the 846 Dosing
Interface.
Touch Control With integrated Touch Control cable
Computer with 6.2151.000 cable
MSB connectors (Metrohm Serial Bus)
Dosing device
Connection for a maximum of 4 external dosing devices, models Dosi-
mat or Dosino (MSB 1 to MSB 4).
Stirrer Connection for a maximum of 4 stirrers.
Rate in 15 steps, shift direction can be selected.
Remote Box Connection for a maximum of four Remote Boxes. Remote Boxes can
be used to actuate and monitor external devices.