8.13 Detector
■■■■■■■■■■■■■■■■■■■■■■
106
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
8.13 Detector
You can find the technical specifications for the detector in the manual for
the detector.
8.14 Power connection
Required supply
voltage
100–240 V ± 10% (autosensing)
Required fre-
quency
50–60 Hz ± 3 Hz (autosensing)
Power consump-
tion
65 W for typical analysis application
25 W standby (conductivity detector to 40 °C)
Power supply unit Up to 300 W maximum, electronically monitored
Internal fuse 3.15 A
8.15 Interfaces
USB
Input 1 USB upstream, type B (labeled PC for connecting to a computer)
Output 2 USB downstream, type A (labeled USB 1 and USB 2)
MSB 2 MSB 8-pin Mini DIN (female) (labeled MSB 1 and MSB 2 for Dosino,
stirrer, remote lines, etc.)
Detector 1 15-pin high-density D-sub (female) (labeled Detector)
Column recogni-
tion
for an intelligent column
Leak sensor 1 jack plug (labeled Leak Sensor)
Further connec-
tions
Auxiliary 1 15-pin DSUB (female)
Service 1 15-pin DSUB (female)

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex ChS/PP/Deg

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex ChS/PP/Deg and is the answer not in the manual?

Metrohm 930 Compact IC Flex ChS/PP/Deg Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex ChS/PP/Deg
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals