EasyManuals Logo

Metrohm 930 Compact IC Flex ChS/PP/Deg User Manual

Metrohm 930 Compact IC Flex ChS/PP/Deg
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
53
Accessories For this step, you need the following accessories:
Guard column (suitable for separation column)
Connecting the guard column
1 2 3
6.2744.070
6.2744.040
6.2744.070
6.2744.070
1
Removing the coupling
Remove the coupling (6.2744.040) installed between the column
inlet capillary and the column outlet capillary for the initial start-up.
2
Preparing the guard column
Remove the stoppers or the stopper and the sealing cap from the
guard column.
3
Connecting the guard column
CAUTION
When inserting the guard column, ensure that it is inserted cor-
rectly based on the marked flow direction (if specified).
Fasten the inlet of the guard column to the column inlet capillary
using a short pressure screw (6.2744.070).
If the guard column is connected to the separation column using
a connection capillary, fasten this connection capillary to the
guard column outlet with a pressure screw.
Rinsing the guard column
1
Rinsing the guard column
Place a beaker under the guard column's outlet.

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex ChS/PP/Deg

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex ChS/PP/Deg and is the answer not in the manual?

Metrohm 930 Compact IC Flex ChS/PP/Deg Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex ChS/PP/Deg
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals