Insert new module for the Autolab 8-Series
25 | Page
Figure 26 The experimentally determined value of C1 is reported in a Message box at the end of
the measurement
9. Open the Hardware setup of Nova (Tools Hardware setup). Select the instrument
type in the Main Module frame in the hardware setup window and adjust the
value of C1 to the value reported in the Message box
4
(see Figure 27).
Figure 27 Change the value of C1 to the value reported in the Message box
4
It is possible to copy and paste the contents of the Message box.