LDM 210 - LDM 220 Corrective Maintenance
6.2 Start-up of Maintenance Modes
6.2.1 REMOTE DOWNLADING AND PROGRAMING Mode
In this mode it is possible to download a new version of the software to the reader over
the RS-232 link for the LDM-210 model or via the USB port for the LDM-220 model. This
new version is stored in the flash memory of the reader DSP chip.
It is necessary to use a remote access software specific to the reader (contact MGP
Instruments for details).
SS1
SS6
J16
Contact
J15, 1-2
To enter this remote programming mode the jumpers should be positioned:
Jumper SS1 (J16),
Jumper SS6 (J16),
One jumper over the contacts 1 and 2 of J15.
Use to download software to a remote PC.
To exit this programming mode you should install :
- jumper SS1,
-jumper SS6,
-jumper over contacts 1 and 2 of J15.