Optical Charts
thermoIMAGER TIM
TIM M-1
TIM M-05
1
(382 x 288 px)
Focal
length
Minimum
distance
2
Distance to object (Measuring field in m, pixel in mm)
0,1 0.2 0.3 0.5 1 2 4 6 10 30 100
f = 16 mm 16 mm 0.2 m HFOV
(m)
0.07 0.11 0.18 0.36 0.72 1.43 2.15 3.6 10.7 35.8
VFOV
(m)
0.05 0.08 0.14 0.27 0.54 1.08 1.62 2.7 8.1 27.0
IFOV
(mm)
0.2 0.3 0.5 0.9 1.9 3.8 5.6 9.4 28.1 93.8
f = 25 mm 25 mm 1.5 m HFOV
(m)
0.023 0.05 0.07 0.11 0.23 0.46 0.92 1.38 2.3 6.9 22.9
VFOV
(m)
0.017 0.03 0.05 0.09 0.17 0.35 0.69 1.04 1.7 5.2 17.3
IFOV
(mm)
0.1 0.1 0.2 0.3 0.6 1.2 2.4 3.6 6.0 18.0 60.0
f = 50 mm 50 mm 1.5 m HFOV
(m)
0.06 0.11 0.23 0.46 0.69 1.1 3.4 11.5
VFOV
(m)
0.04 0.09 0.17 0.35 0.52 0.9 2.6 8.6
IFOV
(mm)
0.2 0.3 0.60 1.2 1.8 3.0 9.0 30.0
HFOV: Horizontal enlargement of the total measuring field at object level
VFOV: Vertical enlargement of the total measuring field at object level
IFOV: Size of the single pixel at object level
1) TIM M-05 is only available with OF25 optics.
2) The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance.