When the memory is accumulated to 99 memory sets.
Under this circumstance, if the device not cleared memory sets ,when the new measurement
is performed, then the new measurement value will always be stored in MR99. i.e. previous
MR99 will be automatically moved to MR98, MR01- the first value will be auto deleted.
Note: How not to store a reading
Press the M-button while the reading is being displayed. Keep the button pressed until <M>
is flashing and then release it.
Confirm by pressing the M-button again.
6. ERROR MESSAGES/MALFUNCTIONS
If an error occurs during a measurement, the measurement is interrupted
and a corresponding error code is displayed.
(Example Error no. 1)
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell
below 20 mmHg. This may occur
, if the tube unplugged after the systolic
blood pressure has been measured.
Further possible cause: No pulse has been detected.
ERR 2 Unnatural pressure impulses influence the measurement result. Reason: The
arm was moved during the measurement (Artefact).
ERR 3 An adequate pressure cannot to generated in the cuff. A leak may have
occurred. Check that the cuff is correctly connected and is not too loose.
ERR 5 The measured reading indicated an unacceptable difference between systolic
and diastolic pressures. Take another measurement following directions care-
fully. Contact your doctor if you continue to get unusual readings.
HI The pressure in the cuff is too high (over 300 mmHg)
OR the pulse is too high (over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40 beats per minute).
Repeat the measurement.*
*Please consult your doctor if this problem occurs repeatedly.
Other possible malfunctions and solutions
If problems occur when using the device, the following points should be checked and if nec-
essary, the corresponding measures are to be taken:
Malfunction Remedy
The display remains empty when the 1. Check the polarity of the batteries (+/-)
device is switched on although
2. If the display is erratic or unusual, remove
batteries have been installed. the batteries and re-install new batteries.
The pressure does not rise although 1. Check the connection of the cuff tube and connect
the pump is running. properly if necessary.
The device frequently fails to 1. Ensure the cuff fits correctly.Please refer to page
measure blood pressure values, or 10 point 5.3
the values measured are too low 2. Make sure that the cuff is not fitted too tightly.
(too high). Ensure tight clothing such as a rolled-up sleeve
isn’t exerting pressure on the arm above the
measuring position. Take off articles of clothing
if necessary.
3. Perform the measurement again in complete
peace and quiet.
Every measurement produces a 1. Please refer to " Further information" listed below.
different value although the
device functions normally and
the values displayed are normal
The Blood pressure values measured 1. Record the daily development of the values and
differ from those measured by the consult your doctor.
doctor. • Please note: Blood pressure fluctuates continually
so successive measurements will show some
variability
Further information
The level of blood-pressure is subject to fluctuations even in healthy people. It is impor-
tant to compare measurements taken under the same conditions and at the same time of
day. (Quiet conditions)!
For licensing, this device has been subjected to strict clinical tests and supervised by
experienced Cardiac Specialists.
This device is manufactured to meet European standards under the supervision of the
15
14
BP 3BM1-3 Automatic Blood Pressure Monitor Instruction Manual www.microlife.com www.microlife.com BP 3BM1-3 Automatic Blood Pressure Monitor Instruction Manual
IB 3BM1-3 for AP EN 2810.qxd 2010.7.16 9:27 AM Page 14

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP 3BM1-3 and is the answer not in the manual?

Microlife BP 3BM1-3 Specifications

General IconGeneral
BrandMicrolife
ModelBP 3BM1-3
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals