MSB780(X) - User’s Guide
Set height difference in m,
ranges -30.00 .. 30.00 : 0.1<cr>
wait..OK<cr><lf>
In the example the new HHCP parameter was set to 0.1 m.
6.49 “hqfe?” Get altitude for QFE corrected pressure
hqfe?<cr>
HQFE: 0.00 m<cr><lf>
6.50 “hqnh?” Get altitude for QNH corrected pressure
hqnh?<cr>
HQNH: 100.00 m<cr><lf>
6.51 “tqfe?” Get default temperature for QFE corrected pressure
This command is used to print the default temperature value used for QFE correction after reset
of the barometer.
Example:
tqfe?<cr>
TQFE: 30.00 C<cr><lf>
6.52 “tqfea?” Get actual temperature for QFE corrected pressure
This command is used to print the temperature value actually being used for QFE correction.
Example:
tqfea?<cr>
TQFE actual: 30.00 C<cr><lf>
6.53 “hhcp?” Get height difference for HCP level
This command is used to print the height difference for HCP corrected pressure.