MSB780(X) - User’s Guide
aXD3!<cr> from transducer 3.
Example:
0XD1!<cr>
0+999.620+10367.5377+0.600085+50.00+0.103364<cr><lf>
Where
0 sdi-12 barometer address
+999.620 pressure in hPa from transducer 1
+10367.5377 frequency in Hz from transducer 1
+0.600085 voltage in V from transducer 1
+50.00 relative humidity
+0.103364 pressure correction from compensation sensor
<cr><lf> line termination
8.15 aXV! Get extended verification
This command return information about firmware version.
Example:
0XV!<cr>
0v. 042,Nov 4 2015,15:45:14<cr><lf>
Where
0 sdi-12 barometer address
v. 042 firmware version
Nov 4 2015 date of build firmware
15:45:14 time of build firmware
<cr><lf> line termination