6–20
MEASUREMENT MODE
:MEAS:FUNC
Select function sub-path within measurement mode.
Parameters:
None.
Response:
None.
:MEAS:FUNC:C, L, X, B, Z, Y, Q, D, R, G
Select first or second AC measurement function.
Selecting first measurement:
:MEAS:FUNC:C Capacitance.
:MEAS:FUNC:L Inductance.
:MEAS:FUNC:X Reactance.
:MEAS:FUNC:B Susceptance.
:MEAS:FUNC:Z Impedance.
:MEAS:FUNC:Y Admittance.
Selecting second measurement:
:MEAS:FUNC:Q Quality factor.
:MEAS:FUNC:D Dissipation factor.
:MEAS:FUNC:R Resistance.
:MEAS:FUNC:G Conductance.
Note that selecting Z or Y as the first measurement will force the second measurement to be
Angle. This does not change the equivalent circuit flag setting.
Example: :MEAS:FUNC:C;D
will select C+D measurements.