7.4 Display Range
7.5 Modes Of Operation
7.5.1 MEASUREMENT
Selection of any measurement parameter and test condition.
Single-level function-menu controlled by keypad and soft keys.
Single and repetitive measurements displaying major and minor terms.
Analogue scale with configurable Hi/Lo limits giving PASS/FAIL indication.
7.5.2 MULTI-STEP
Measurement parameters and test conditions set using MULTI-STEP SET MODE.
Up to 30 steps with configurable limits.
PASS/FAIL indication.
Up to 64 multi-step programs can be saved in the non-volatile memory.
7.6 Measurement Connections
4 front panel BNC connectors with the screens at ground potential.
Terminals withstand connection of charged capacitor up to 5V, either polarity.
7.7 Measurement Accuracy
The accuracy statements given apply when the instrument is used under the following
measurement conditions.