6–26
MULTI-STEP MODE
:MULTI:TEST
Select the step to edit.
Parameters:
The step number in the range 1 to 30.
Example: :MULTI:TEST 1
will select the top frequency for editing
Response:
None.
:MULTI:TEST?
Return the number of the step that is
currently being edited.
Parameters:
None.
Response:
The frequency number in the range 1 to
30.
Example: 30
would indicate the last frequency is
selected for editing.
:MULTI:FUNC
Set measurement function for currently
selected step.
Parameters:
The following discrete parameters are
valid:
LS, LP, Q, CS, CP, D, Z, PHASE,
RS, RP, X, G, B, Y, RDC.
Response:
None.
:MULTI:FUNC?
Parameters:
None.
Response:
Returns the measurement type according to
This table:
1 Z 9 P
2 Ls 10 Q
3 Lp 11 D
4 Cs 12 Rs
5 Cp 13 Rp
7 Y 14 B
8 G 15 X
16 RDC