6–30
MULTI-STEP MODE
:MULTI:TRIGger
Starts a run of multi-step measurements.
Parameters:
None.
Response:
None.
:MULTI:RES?
Query the results of the multi-step test.
Parameters:
None
Response:
Each result is separated by a comma and the PASS/FAIL flag will prefix the result.
0 PASS,
1 FAIL (HI)
2 FAIL (LO)
Examples: 1, +.68898363E-07, 0, +.72168059E-04
would indicate a fail result for step 1 and a pass result for step 2.
MULTI:NEW <filename>
Create a file under a new name to store the multi-step data.
Parameters:
A File name, 8 characters at most.
Example: :MULTI:NEW DEMO
will create a new file with the name DEMO.
Response:
None.