Amicon
®
Ultra-2 Centrifugal Filter Devices
1
User Guide
Amicon
®
Ultra-2
Centrifugal Filter Devices
for volumes up to 2 mL
For research use only;
not for use in diagnostic procedures.
Introduction
Amicon
®
Ultra-2 centrifugal lter devices provide fast ultraltration, with the capability for high concentration factors and easy
concentrate recovery from dilute and complex sample matrices. The vertical design and available membrane surface area provide
fast sample processing, high sample recovery (typically greater than 90% of dilute starting solution), and the capability for
50-fold concentration. Typical processing time is 10 to 60 minutes depending on Molecular Weight Cut O (MWCO). Solute
polarization and subsequent fouling of the membrane are minimized by the vertical design, and a physical deadstop in the lter
device prevents spinning to dryness and potential sample loss. Ecient recovery of the concentrated sample (retained species) is
achieved by a convenient reverse spin step after collecting the ltrate. The device can be spun in a swinging bucket or xed angle
rotor. Amicon
®
Ultra-2 devices are supplied non-sterile and are for single use only.
The Amicon
®
Ultra-2 product line includes 5 dierent cutos (Molecular Weight Cut O, MWCO). These devices are for
research use only and not for use in diagnostic procedures.
• Amicon
®
Ultra 3K device — 3,000 MWCO
• Amicon
®
Ultra 10K device — 10,000 MWCO
• Amicon
®
Ultra 30K device — 30,000 MWCO
• Amicon
®
Ultra 50K device — 50,000 MWCO
• Amicon
®
Ultra 100K device — 100,000 MWCO
Applications
• Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids (DNA/RNA samples, either
single- or double-stranded), microorganisms, column eluates, and puried samples
• Purication of macromolecular components found in tissue culture extracts and cell lysates, removal of primer, linkers, or
molecular labels from a reaction mix, and protein removal prior to HPLC
• Desalting, buer exchange, or dialtration
Materials Supplied
10K
4
3
2
1
00
20
00
00
00
Filter Device
Concentrate
collection
tube
Filtrate
collection
tube
The Amicon
®
Ultra-2 device is supplied with two tubes. During
operation, one tube is used to collect ltrate; the other to cap
the device during concentration and subsequently to recover the
concentrated sample.