MAINTENANCE
85
Section 5-2 Replacing the Q-Gard Pack
When
The Q-Gard Pack should be replaced when one of the following Alert or Alarm
messages is displayed.
x Alert message = REPLACE Q-GARD PACK, or
x Alarm message = MILLI-Q RES < SP, REPLACE Q-GARD AND
QUANTUM
Removal
procedure
Follow the steps below to remove the Q-Gard Pack.
Step Action Result
1 Make sure the Milli-Q System is
in STANDBY Mode.
2 While in STANDBY Mode, push the Q-POD Dispenser Plunger down
once to depressurise the Milli-Q System. After water stops being
dispensed, push down the Q-POD Dispenser Plunger again.
Continued on next page