31
Alarms generated during installation
Overview
• During the installation of a Milli-Q System, certain Alarm messages are
generated.
• This occurs because:
− the Reservoir is empty,
− there is air in the tubings and in the Progard Pack,
− the Progard Pack is not installed,
− the Quantum Cartridge is not installed, and
− an automatic flush of the Progard Pack is manually cancelled.
• These alarms are explained here. The ways to cancel them are explained
also. For more information about Alarm messages, see the chapter titled
‘Alarms’.
• It is perfectly normal to see alarms during installation.
• The Milli-Q System is designed to use various sensors to alert you of
problems during normal operation of the system. This insures optimal
water quality.
• During installation, these sensors are active. As a result, it is possible to
have alarms generated. In order to advance during the installation, these
alarms should be cancelled.
TANK EMPTY
message
• This alarm occurs because the Reservoir is empty during most of the
installation.
• This alarm goes away when the Reservoir is partially full.
• To cancel the text display of this alarm message, follow the instructions on
the LCD.
PROGARD
PACK OUT
message
• This alarm occurs because the Progard Pack is not installed.
• This alarm goes away when the Progard Pack is detected by the Milli-Q
System.
• To cancel the text display of this alarm message, follow the instructions on
the LCD.
QUANTUM
CARTRIDGE
OUT message
• This alarm occurs because the Quantum Cartridge is not installed.
• This alarm goes away when the Quantum Cartridge is detected by the
Milli-Q System.
• To cancel the text display of this alarm message, follow the instructions on
the LCD.