Page 8 XRS-5 Transducer PL-590
About this
General Guidelines
The equipment may be used in all hazardous zones with all gases with
temperature classes T1, T2, T3, T4, T5 and T6. The equipment is only
certified for use in ambient temperatures in the range of -20°C to +65°C and
should not be used outside this range.
Installation shall be carried out in accordance with the applicable code of
practice by suitably trained personnel.
The apparatus shall only be supplied from a circuit containing a suitably-
rated fuse having a breaking capacity of 4000A. This fuse is included in
Milltronics’ transceivers.
Repair of this equipment shall be carried out in accordance with the
applicable code of practice.
The certification of this equipment relies on the following materials used in its
construction:
Enclosure: Kynar Flex 2800-02 (former designation 2820) /
Chlorosulfonated polyethylene / Nitrile / Ethylene propylene
/ Chloroprene
Encapsulant: LA-9823-76
Manual override can be accomplished by using the disconnect switch
provided in the building installation of the associated controller.