3-3
NOTE
Make sure the pads package is intact before use. Otherwise, replace it with a new one.
3.2 Basic Operations
3.2.1 Turning Power On
1. Check for mechanical damages on the equipment or other damages on the pads package.
2. Make sure the pads cable is properly connected and battery installed.
3. Check the expiration date of the pads on the pads package.
4. Press the Power ON/OFF button to start the equipment.
3.2.2 Changing General Settings
You can change the general settings in the [General Setup] menu.
To access the [General Setup] menu,
If the equipment is on, press the Power ON/OFF button and the “Select an option” window is displayed.
Then .select
[Config.][Config. Edit]enter the required password[General Setup].
If the equipment is off:
For Pro, press the Power ON/OFF button, the third and fourth soft keys (from left to right) simultaneously to
display the maintenance screen;
For Public, press the Power ON/OFF button, the second and third soft keys (from left to right)
simultaneously.
Then select [Config.][Config. Edit]enter the required password[General Setup].
NOTE
All changes made in Configuration mode are auto-saved immediately. You can turn off the equipment
after the setting is finished.
3.2.2.1 Setting the Date and Time
1. In the [General Setup] menu, select [System Date] to set the system date.
2. Select [Time] to set the system time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D1 and is the answer not in the manual?

Mindray BeneHeart D1 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D1
CategoryMedical Equipment
LanguageEnglish

Related product manuals