9-4
9.6.4 CPR Setup Menu
Menu Item Options/Range Default
CPR Mode 30:2, 15:2, Hands-Only 30:2
9.6.5 ECG Setup Menu (For Pro Only)
Menu Item Options/Range Default
ECG Bandwidth Monitor, Therapy Therapy
Arrhythmia Lethal Arrh. Only, All Lethal Arrh. Only
Asystole Delay 3 to 10 5
Extreme Tachy
Adult 60 to 300 160
Pediatric 60 to 300 180
Extreme Brady
Adult 15 to 120 35
Pediatric 15 to 120 50
Vbrd Rate 15 to 60 40
Vbrd PVCs 3 to 99 5
V-Tach Rate 100 to 200 130
V-tach PVCs 3 to 99 6
Tachy
Adult 60 to 300 120
Pediatric 60 to 300 160
Brady
Adult 15 to 120 50
Pediatric 15 to 120 75
Multif. PVCs Window 3 to 31 15
PVCs High 1 to 100 10
ECG Standard AHA, IEC AHA
Pacemaker Detection On, Off Off
9.6.6 Alarm Setup Menu (For Pro Only)
Menu Item Options/Range Default
Alarm Volume High, Med, Low, Off Low
Reminder Tone On, Off Off
9.6.7 Test Setup Menu
Menu Item Options/Range Default
User Test Setup 00:00, 01:00, 02:00, 03:00, 04:00, 05:00, 03:00,