A-3
Selected energy accuracy
Impedance
Energy
25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Accuracy
1 J 1 1 1 0.9 0.9 0.9 0.8 ±2J
2 J 2 2 2 1.9 1.8 1.7 1.6 ±2J
3 J 2.9 3 2.9 2.8 2.7 2.6 2.4 ±2J
4 J 3.9 4 3.9 3.7 3.6 3.4 3.2 ±2J
5 J 4.9 5 4.9 4.7 4.5 4.3 4.1 ±2J
6 J 5.8 6 5.8 5.6 5.3 5.1 4.9 ±2J
7 J 6.8 7 6.8 6.6 6.3 6 5.7 ±2J
8 J 7.8 8 7.8 7.4 7.1 6.8 6.5 ±2J
9 J 8.8 9 8.8 8.4 8 7.7 7.3 ±2J
10 J 9.7 10 9.7 9.3 8.9 8.5 8.1 ±2J
15 J 15 15 15 14 13 13 12 ±15%
20 J 20 20 20 19 18 17 16 ±15%
30 J 29 30 29 28 27 25 24 ±15%
50 J 49 50 49 47 45 43 41 ±15%
70 J 68 70 68 65 62 60 57 ±15%
100 J 97 100 97 93 89 85 81 ±15%
150 J 146 150 146 140 134 128 122 ±15%
170 J 166 170 166 159 151 145 138 ±15%
200 J 195 200 195 187 178 170 163 ±15%
300 J 292 300 292 280 267 255 244 ±15%
360 J 351 360 350 336 321 306 293 ±15%
Charge time (Note: at 20 C of ambient temperature)
Charge time
(Manual Defib)
From initiation of rhythm
analysis to charge done
From initial power
on to charge done
200J 360J 200J 360J 200J 360J
Rechargeabl
e battery
New and fully charged <5 s <8 s <10 s <10 s <20 s <20 s
New and fully charged after
15 times of 360J discharges
<6 s <9 s <10 s <10 s <20 s <20 s
Disposable
battery
New / / <10 s <17 s <20 s <27 s
New after 15 times of 360J
discharges
/ / <10 s <17 s <20 s <27 s
Synchronized discharge delay
Local synchronized discharge delay < 60ms (from the peak of R-wave)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D1 and is the answer not in the manual?

Mindray BeneHeart D1 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D1
CategoryMedical Equipment
LanguageEnglish

Related product manuals