Defibrillator/Monitor Operator’s Manual 15 - 1
15 Marking Events
During patient monitoring or therapy, some events may exert effects on the patient and as a result change
related waveforms and parameter values. To help analyzing the waveforms or numerics at that time, you can
mark these events.
Before marking an event, you can define events A to F, for example, define event D as injecting Atropine. You can
only define events by selecting [Mark Event Setup] from the Configuration Main menu. Event A is always
[Generic] and cannot be changed.
To mark an event,
1. In the Monitor mode, Manual Defib, mode or Pacer mode, press the Event button on the front panel to
enter the [Mark Event] menu.
2. Select an event you want to mark from [A] to [H], or select [Exit] to return to the main screen.
In the AED mode, pressing the Event button on the front panel directly marks Event A “Generic”.
When you mark an event, the event name and the time when the event is triggered will be displayed at the
prompt information area. This information disappears automatically after a period of 5 seconds.