Defibrillator/Monitor Operator’s Manual D - 1
D Defibrillator Shift Checklist
Equipment Name: Serial Number: Department:
Item Pass/
Fail
Corrective Actions/Remarks
1. Equipment Appearance
Clean, no foreign substance, no crack
2. Cables/Connectors
Cables not frayed, connectors and pins not broken or
loose
Connectors engage securely
3. Batteries
Battery installed with at least 60% of battery capacity
Fully charged spare battery available
4. Necessary Accessories (Electrode pads, electrodes,
paddles or recorder paper)
Present and sufficient
Inspected to be used normally
5. Auto Test
Status indicator off
6. Shock Test
If the external paddles are used:
a. Connect the equipment with AC power supply, and
the AC power indicator is illuminated.
b. Connect the paddles cable to the equipment, and
place the external paddles in the paddle tray.
c. Press the Charge button on the external paddles, and
charge the equipment to 50J.
d. Press the Shock button on the external paddles.
e. The system prompts that the shock is delivered
normally.
6. Shock Test
If the electrode pads are used:
a. Connect the equipment with AC power supply, and
the AC power indicator is illuminated.
b. Connect the pads cable to the equipment.
c. Perform the Energy Delivery test item of the user test
with the test load connected.
d. The system prompts that the Energy Delivery test is
passed.
7. Monthly Check on Expiration Date
The electrode pads are not expired.
Checked by:
Date: