21 - 6 Defibrillator/Monitor Operator’s Manual
Lead Set 3-Lead, 5-Lead 3-Lead Sets the ECG lead type.
This setting affects the default
waveform position of the
[Waveform Setup] menu.
QRS Volume 0 to 10 2 This setting is linked with the
[QRS Volume] setting in the
[SpO2 Setup] menu
ECG1 3-Lead: I, II, III II Available options are defined by
the current setting of [Lead
Set].
5-Lead: I, II, III, aVL, aVR, aVF,
V
ECG2 3-Lead: I, II, III / • If 3-lead is used, the default
setting of this item is blank
and this item is inactivate.
• The options of [ECG1] are
not available for [ECG2].
5-Lead: I, II, III, aVL, aVR, aVF,
V
V
Sweep 6.25 mm/s, 12.5 mm/s, 25
mm/s, 50 mm/s
25 mm/s Sets the ECG waveform speed.
HR Alarm On, Off On /
HR Alm Lev High, Med Med /
HR High Adu (Low+2) to 300 120 Step: 1 bpm
Ped (Low+2) to 300 160
Neo 200
HR Low Adu 15 to (High-2) 50
Ped 75
Neo 100
Arrhythmia On, Off Off Sets whether to enable the
arrhythmia analysis.
ARR Alarm On, Off On /
Menu Item Options/Range Default Description