Defibrillator/Monitor Operator’s Manual 4 - 5
NOTE
• You can enable auto alarm limits only when the current parameter measurement is within the auto
alarm limits range.
Module Parameter Low alarm limit High alarm limit Auto alarm limits range
Adult/pediatric Neonate Adult/pediatric Neonate
ECG HR HR×0.8 or 40
bpm (whichever
is greater)
(HR-30) or 90
bpm (whichever
is greater)
HR×1.25 or 240
bpm whichever
is lower)
(HR + 40) or 200
bpm whichever
is lower)
Adult/pediatric: 35 to 240
Neonate: 55 to 225
Resp RR RR×0.5 or 6/min
(whichever is
greater)
(RR-10) or 30/
min (whichever
is greater)
RR×1.5 or 30/
min (whichever
is lower)
(RR+25) or 85/
min (whichever
is lower)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
SpO
2
SpO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
PR PR×0.8 or 40
bpm (whichever
is greater)
(PR-30) or 90
bpm (whichever
is greater)
PR×1.25 or 240
bpm whichever
is lower)
(PR + 40) or 200
bpm whichever
is lower)
Adult/pediatric: 35 to 240
Neonate: 55 to 225
NIBP NIBP-S SYS×0.68 +
10mmHg
(SYS-15) or
45mmHg
(whichever is
greater)
SYS×0.86 +
38mmHg
(SYS + 15) or
105mmHg
(whichever is
lower)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
NIBP-D Dia×0.68 +
6mmHg
(Dia-15) or
20mmHg
(whichever is
greater)
Dia×0.86 +
32mmHg
(Dia + 15) or
80mmHg
(whichever is
lower)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
NIBP-M Mean×0.68 +
8mmHg
(Mean-15) or
35mmHg
(whichever is
greater)
Mean×0.86 +
35mmHg
(Mean + 15) or
95mmHg
(whichever is
lower)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to 105
CO
2
EtCO
2
0-32mmHg:
remains
unchanged
0-32mmHg:
remains
unchanged
0-32mmHg:
remains
unchanged
0-32mmHg:
remains
unchanged
Same as the
measurement range
32-35mmHg:
29mmHg
32-35mmHg:
29mmHg
32-35mmHg:
41mmHg
32-35mmHg:
41mmHg
35-45mmHg:
EtCO
2
-6mmHg
35-45mmHg:
EtCO
2
-6mmHg
35-45mmHg:
EtCO
2
+6mmHg
35-45mmHg:
EtCO
2
+6mmHg
45-48mmHg:
39mmHg
45-48mmHg:
39mmHg
45-48mmHg:
51mmHg
45-48mmHg:
51mmHg
>48mmHg:
remains
unchanged
>48mmHg:
remains
unchanged
>48mmHg:
remains
unchanged
>48mmHg:
remains
unchanged
FiCO
2
N/A N/A Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR awRR×0.5 or 6/
min (whichever
is greater)
awRR-10 or 30/
min (whichever
is greater)
awRR×1.5 or 30/
min (whichever
is lower)
awRR+25 or 85/
min (whichever
is lower)
Adult/pediatric: 6 to 55
Neonate: 10 to 90

Table of Contents

Other manuals for Mindray BeneHeart D2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D2 and is the answer not in the manual?

Mindray BeneHeart D2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D2
CategoryMedical Equipment
LanguageEnglish

Related product manuals