A - 3
A.6 Interface Specifications
A.7 Battery Specifications
USB connector 1, USB 2.0
micro USB connector 1, supports Windows 7 or above operating system
Network connector 1, connects the Wi-Fi or cellular (2G/3G/4G) network.
Multifunction connector 1, connects the CPR sensor.
Battery type Disposable battery
Battery voltage 12V
Battery capacity 4200mAh
Equipment configured without 
the screen
Operating 
time
Testing condition
≥ 15 hours The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, wireless function off, not 
performing defibrillation charges or discharges, voice 
volume set to low.
300 200J 
discharges
The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, wireless function off, voice 
volume set to low, with one minute of CPR between 
discharges
190 360J 
discharges
510 150 J 
discharges
The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, wireless function off, voice 
volume set to low, with three discharges every minute
400 200J 
discharges
200 360J 
discharges
Equipment configured with the 
screen
≥ 12 hours The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, wireless function off, not 
performing defibrillation charges or discharges, voice 
volume set to low.
270 200J 
discharges
The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, wireless function off, voice 
volume set to low, with one minute of CPR between 
discharges
170 360J 
discharges
450 150J 
discharges
The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, wireless function off, voice 
volume set to low, with three discharges every minute
350 200J 
discharges
200 360J 
discharges
Battery fuel gauge (for 
equipment configured with the 
screen)
Battery symbol on the display indicating the current battery level