C - 1
C EMC and Radio Regulatory Compliance
C.1 EMC
The equipment meets the requirements of IEC 60601-1-2: 2014.
• Use of accessories, transducers and cables other than those specified or provided by the 
manufacturer of this device could result in increased electromagnetic emissions or decreased 
electromagnetic immunity of this device and result in improper operation.
• Use of this device adjacent to or stacked with other device should be avoided because it could result 
in improper operation. If such use is necessary, this device and the other device should be observed 
to verify that they are operating normally.
• Portable RF communications equipment (including peripherals such as antenna cables and external 
antennas) should be used no closer than 30 cm (12 inches) to any part of the this device, including 
cables specified by the manufacturer. Otherwise, degradation of the performance of this device 
could result.
• Other devices may affect this equipment even though they meet the requirements of CISPR.
• When the inputted signal is below the minimum amplitude provided in technical specifications, 
erroneous measurements could result.
• The equipment needs special precautions regarding EMC and needs to be installed and put into 
service according to the EMC information provided below.
• Portable and mobile RF communications equipment may affect this equipment.
• This equipment is intended for use in professional healthcare facility environment, or in home 
healthcare environment such as restaurants, cafes, shops, stores, markets, schools, churches, 
libraries, outdoors (streets, sidewalks, parks), domiciles (residences, homes, nursing homes), train 
stations, bus stations, airports, hotels, hostels, pensions, museums, theatres. If it is used in special 
environment, such as magnetic resonance imaging environment, the equipment may be disrupted 
by the operation of nearby equipment.
Guidance and Declaration - Electromagnetic Emissions
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the 
user of the equipment should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The equipment uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to cause 
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The equipment is suitable for use in all establishments, including 
domestic establishments and those directly connected to the 
public low-voltage power supply network that supplies buildings 
used for domestic purposes.