A - 4      
• If the equipment is connected through the wireless network in low strength signal, the battery 
standby life will be shortened.
A.8 Data Storage
Remaining charge after “Low 
Battery” is prompted
For BeneHeart C1/BeneHeart C1A/BeneHeart C1 Fully Automatic/
BeneHeart C1A Fully Automatic/BeneHeart S1/BeneHeart S1A/BeneHeart 
S1 Fully Automatic/BeneHeart S1A Fully Automatic:
• At least 30 minutes operating time (at 20 °C± 5 °C of ambient 
temperature, wireless function off, voice volume set to low) and at 
least 10 200J discharges (with one minute of CPR between discharges) 
• At least 30 minutes operating time (at 20 °C± 5 °C of ambient 
temperature, wireless function off, voice volume set to low) and at 
least 6 360J discharges (with one minute of CPR between discharges) 
For BeneHeart C2/BeneHeart C2A/BeneHeart C2 Fully Automatic/
BeneHeart C2A Fully Automatic/BeneHeart S2/BeneHeart S2A/BeneHeart 
S2 Fully Automatic/BeneHeart S2A Fully Automatic:
• At least 30 minutes operating time (at 20 °C± 5 °C of ambient 
temperature, wireless function off, voice volume set to low, screen 
brightness set to indoor mode) and at least 10 200J discharges (with 
one minute of CPR between discharges)  
• At least 30 minutes operating time (at 20 °C± 5 °C of ambient 
temperature, wireless function off, voice volume set to low, screen 
brightness set to indoor mode) and at least 6 360J discharges (with 
one minute of CPR between discharges) 
Battery standby life 
Standby life Testing condition
5 years  The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, performing auto test every 
day, equipment not in use, not sending selftest report
3 years The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, performing auto test every 
day, equipment not in use, sending selftest report 
every week through the wireless network
2 years  The equipment is powered by a new battery at 20 °C± 5 
°C of ambient temperature, performing auto test every 
day, equipment not in use, sending selftest report 
every day through the wireless network
Waveform storage Up to 5 hours of ECG waveforms
Events Up to 500 events
Voice recording Up to 1 hour
CPR data Up to 5 hours
Selftest reports 1000 records