E - 1
E Voice Prompts
The following table lists voice prompts that may occur during a rescue. 
Condition Voice Prompt Description
Open the lid Powered on. Stay calm. Follow the 
instructions.
The lid is opened.
Device error. Recommended to replace 
the Device. Stay calm. Follow the 
instructions.
The equipment malfunctions, use 
one standby equipment or start CPR 
immediately.
After turning on the 
equipment
Adult mode The Adult/Child mode switch is 
pressed to Adult, or the electrode 
pads connected to the equipment 
are detected for the adult patient.
Child mode. if the patient is an Adult, 
adjust the Adult/Child mode switch to 
Adult mode.
The Adult/Child mode switch is 
pressed to Child.
Child mode The Adult/Child mode switch is 
pressed to Child, or the electrode 
pads connected to the equipment 
are detected for the children.
Place the electrode 
pads
Remove clothing from patient's chest. 
Apply pads as shown on Pads.
Prompts to remove the patient’s 
clothing and place the electrode 
pads.
Remove clothing from patient's chest. 
Plug in pads connector.
Remove pads package from lid of 
AED.Tear open package. Apply pads as 
shown on Pads. 
Apply pads as shown on Pads.
Apply pads as shown on Pads.
Abnormal Pads connection. Pads connection failure, start CPR 
immediately.
The equipment 
analyzes the patient’s 
heart rhythm. 
Do not touch the patient. Analyzing heart 
rhythm. 
Repeats until analysis of the 
patient’s heart rhythm is completed. 
This prompt will be interrupted if 
the equipment is ready to shock.
No shock advised. Notifies non-shockable rhythm has 
been detected.
Motion detected. Do not touch or move 
the patient.
The equipment detects ECG noise 
artifacts, stop moving or touching 
the patient. 
Noise detected. Make sure pads are firmly 
attached.
The equipment detects ECG noise 
artifacts, better pads contact on the 
patient’s skin is required. 
Pads off. Analysis interrupted. Pads connection failure, the 
equipment automatically stops the 
heart rhythm analysis. Reconnect 
the electrode pads.