D - 2      
D.2 AED Setup
D.3 CPR Setup
Menu Item Description Options/Range Default
Shock Series Sets the number of shocks.
If it is set to greater than one, 
the equipment resumes 
analyzing the patient’s rhythm 
after the shock is delivered to 
determine if the shock was 
successful. Prompts for shock 
counter are provided to guide 
you delivering additional 
shocks. 
1, 2, 3 1
Energy 1 (Adult) Sets the defibrillation energy 
level for the first shock on the 
adult patient. 
100, 150, 170, 200, 300, 360J 200 J
Energy 2 (Adult) Energy 1 ≤ configurable value 
≤ Energy 3
Energy 1 to 360J 300 J
Energy 3 (Adult) Energy 2 ≤ configurable value Energy 2 to 360J 360 J
Energy 1 (Pediatric) Sets the defibrillation energy 
level for the first shock on the 
children. 
10, 15, 20, 30, 50, 70, 100J 50 J
Energy 2 (Pediatric) Energy 1 ≤ configurable value 
≤ Energy 3
Energy 1 to 100 J 70 J
Energy 3 (Pediatric) Energy 2 ≤ configurable value Energy 2 to 100 J 100 J
Initial CPR Selects whether the 
equipment enters the CPR 
status directly after turned on.
On, Off Off
ECG Display Selects whether the ECG 
waveform is displayed. 
On, Off Off
Auto Release Time Sets the time the equipment 
automatically removes the 
stored energy internally.
30s, 60s, 90s, 120s 30s
Menu Item Description Options/Range Default
CPR Mode (Adult) Sets the rate of compression 
and ventilation. 
30:2, 15:2, Hands-Only 30:2
CPR Mode (Pediatric) 15:2
CPR Voice Prompts Selects whether voice prompts 
are provided when using a CPR 
metronome. 
On, Off On