C - 2      
If the device is operated within the electromagnetic environment listed in Table Guidance and Declaration - 
Electromagnetic Immunity, the equipment will remain safe and provide the following essential performance: 
energy accuracy, CPR function, data stored.
Guidance and Declaration - Electromagnetic Immunity 
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the 
user of the equipment should assure that it is used in such an environment.
Immunity test IEC 60601 test 
level
Compliance 
level
Electromagnetic environment - guidance
Electrostatic 
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic 
tile. If floors are covered with synthetic 
material, the relative humidity should be at 
least 30%.
Power frequency 
(50/60 Hz) 
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic fields should be at 
levels characteristic of a typical location in a 
typical commercial or hospital environment.
Guidance and Declaration - Electromagnetic Immunity 
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the 
user of the equipment should assure that it is used in such an environment.
Immunity test IEC 60601 Test level Compliance 
level
Electromagnetic environment - guidance
Conduced RF
IEC 61000-4-6
3 Vrms
150k to 80 MHz
3 Vrms (V1) Portable and mobile RF communications 
equipment should be used no closer to any 
part of the device, including cables, than the 
recommended separation distance calculated 
from the equation applicable to the frequency 
of the transmitter. Recommended separation 
distance:
 150k to 80 MHz
 80 MHz to 800 MHz
 800 MHz to 2.7 GHz
where P is the maximum output power rating 
of the transmitter in watts (W) according to 
the transmitter manufacturer and d is the 
recommended separation distance in meters 
(m)
b
.
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site 
survey
c
, should be less than the compliance 
level in each frequency range 
d
.
Interference may occur in the vicinity of 
equipment marked with the following 
symbol: 
6 Vrms
in ISM bands and 
amateur radio bands
a
 
between 0.15 MHz 
and 80 MHz
6 Vrms (V2)
Radiated RF EM 
fields
IEC 61000-4-3
3V/m
80 MHz to 2.7 GHz
3 V/m (E1)
10V/m
80 MHz to 2.7 GHz 
(IEC60601-2-4)
10 V/m
20V/m
80 MHz to 2.7 GHz 
(IEC60601-2-4)
20 V/m
Proximity fields 
from RF wireless 
communication
s equipment
IEC61000-4-3
27 V/m
380 to 390 MHz
27 V/m
28 V/m
430 to 470 MHz, 
800 to 960 MHz, 1700 
to 1990 MHz, 2400 to 
2570 MHz
28 V/m
9 V/m
704 to 787 MHz, 5100 
to 5800 MHz
9 V/m