Image Optimization 5-59
Angle
The range from acquisition beginning to acquisition end.
Range: 1075°.
5.10.6.3 Color STIC 3D
The system also supports color STIC 3D flow image function. For details, please refer to 5.10.10
Color 3D.
5.10.7 3D Layout
The function compiles the 3 MPRs together according to their relative positions, which provides a
much clearer interior anatomical structure.
This function is provided by [Niche] option, and it does not support Smart 3D imaging.
Only probe D7-2E, D8-2E, D6-2NE, DE11-3E and DE10-3E support Niche function.
Niche
1. Tap [Tools][3D Layout] on the touch screen, and then touch [Niche]. .
2. Touch [A]/[B]/[C]/[Niche] on the touch screen to select the reference plane as Plane A, Plane B,
Plane C or Niche.
3. Set the view direction for niche display mode using touch screen: from the front of the
reference image or from the back.
4. Roll the trackball to view sectional images as necessary; you can rotate <M>, <PW>, <Color>
to perform axial rotation or rotate <4D> knob to adjust the nearest VOI section (cut plane)
position.
Click [Reset Ori] to reset the rotation, the center point position, as well as the position of the
MPR.
3Slice
1. Tap [Tools][3D Layout] tab on the touch screen, and then touch [3Slice]. .
2. Touch [A]/[B]/[C]/[3Slice] on the touch screen to select the reference plane.
3. Roll the trackball to view sectional images as necessary; you can rotate <M>, <PW>, <Color>
to perform axial rotation or rotate <4D> knob to adjust the nearest VOI section (cut plane)
position.
Click [Reset Ori] to reset the rotation, the center point position, as well as the position of the
MPR.