11-8 DICOM/HL7
Select an item in the service list, change the parameters in the above area, and click [Update]
to update the item in the service list; click [Cancel] to cancel changing.
Select an item in the service list, and click [Delete] to delete the service.
Select an item in the service list, and click [Default] to set the server to be the default service.
HL7 service setting for Worklist is described as follows:
Name
NOTE
Configure
New service
Device Name
After you set the server (s) in DICOM Server Setting
screen, the name (s) will appear in the drop-down list,
select the name of the Worklist server.
DICOM Service Name
Default is server-HL7Query, and it can be modified.
AE Title
Application Entity title.here, it should be consistent with
that of the HL7 server.
Port
DICOM communication port, 104 by default.Here, the
port should be consistent with that of the HL7 server
port.
Maximum Retries
Reserved feature.
Interval Time(s)
Reserved feature.
Timeout
Refers to time after which the system will stop trying to
establish a connection to the service.Value: 560s, in
increments of 5, and 15 by default.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 EXP and is the answer not in the manual?

Mindray DC-80 EXP Specifications

General IconGeneral
TouchscreenYes
Transducer Ports4
Power Supply100-240V AC, 50/60 Hz
Imaging ModesColor Doppler, Power Doppler
Operating SystemWindows
Storage1 TB HDD
ConnectivityDICOM, Ethernet, USB
ApplicationsGeneral Imaging, Cardiology, Vascular, Pediatrics
Weight85 kg