S-6 
 10.Before examining a new patient, press 『Patient』 to delete the 
patient information and data recorded in the memory for the 
previous patient. Otherwise, the new data may be confused with the 
data of the previous patient. 
    11. Do not pull out plugs of the system and its accessories without 
turning OFF the power. Otherwise it may result in equipment damage 
or even electric shock. 
    12. Do not turn OFF the power supply of the system during printing, 
saving, or invoking. Otherwise it may result in abnormality of these 
processes. 
 
NOTE:  1.  Do not use the system in the vicinity of strong electromagnetic field (such as 
the transformer), which may affect the performance of the monitor. 
  2.  Do not use the system in the vicinity of high-frequency radiation source 
(such as the cellular phone), which may affect the performance of the 
machine or even lead to failure. 
  3. To avoid damaging the system, do not use the system in following 
environment:  
      (1) Locations exposed to direct sunlight 
      (2) Locations subject to sudden changes in temperature 
  (3) Dusty locations 
      (4) Locations subject to vibration 
      (5) Locations near heat generators 
      (6) Locations with high humidity 
  4.  Turn ON the system only after the power has been OFF for more than 10 
seconds. If the system is turned ON immediately after being turned OFF, it 
may result in malfunction of the system. 
  5. Turn OFF the auxiliary power switch or stop transmission through the 
[Freeze] key before connecting or disconnecting a transducer. If a 
transducer is connected or disconnected with an image displayed, the 
system and/or the transducer may malfunction. 
  6.  After using the transducer, remove the ultrasound gel on it and place the 
transducer on the transducer holder. Otherwise, water in the gel may enter 
the acoustic lens, thus adversely affecting the performance and safety of the 
transducer.