Electrical Safety Inspection
A-3
ELECTRICAL SAFETY INSPECTION
2- Device Enclosure And Accessories
TEST PROCEDURE
Visual Inspection
TheEnclosureandAccessories
No physical damage to the enclosure and
accessories.
No physical damage to meters, switches,
connectors,etc.
No residueof fluid spillage (e.g., water, coff ee,
chemicals,etc.).
No loose or missing parts (e.g., knobs, dials,
terminals,etc.).
Contextual Inspection
TheEnclosureandAccessories
No unusual noises (e.g., a rattle inside the
case).
No unusual smells (e.g., burning or smoky
smells,particularlyfromventilationholes).
No taped notes that may suggest device
deficienciesoroperatorconcerns.