Electrical Safety Inspection
A-2
ELECTRICAL SAFETY INSPECTION
1- Power Cord Plug
TEST PROCEDURE
The Power Plug
ThePowerPlugPins Nobrokenorbentpin.Nodiscoloredpins.
ThePlugBody Nophysicaldamagetotheplugbody.
TheStrainRelief
Nophysicaldamagetothestrainrelief.Noplug
warmthfordeviceinuse.
ThePowerPlug Nolooseconnections.
The Power Cord
ThePowerCord
No physical damage to the cord.No
deteriorationtothecord.
‐‐For devices with detachable power cords,
inspecttheconnectionatthedevice.
‐‐Fordeviceswithnondetachablepowercords,
inspectthestrainreliefatthedevice.