EasyManuals Logo

Mindray ME7 Service Manual

Mindray ME7
292 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #288 background imageLoading...
Page #288 background image
Part III Appendix
Electrical Safety Maintenance 271
Earth leakage
current
Single fault condition
(SFC)
____μA
NC: 500 μA
SFC: 1000 μA
6*
Patient leakage
current
Normal condition (NC)
BF____μA
Limit
BF application type
NC: 100 μA
SFC: 500 μA
Single fault condition
(SFC)
BF____μA
7* Leakage current on the major applied part
BF____μA
Limit
BF application type
5000 μA
8*
Patient auxiliary
current
Normal condition (NC)
BF____μA
Limit
NC:10 μA
SFC: 50 μA
Single fault condition
(SFC)
BF____μA
Note: The 4-protection ground impedance test item is applicable only to the MT3 trolley (with power) and not to
the standalone MX portable device or the MT2 trolley (without power).
Name: _____________________________Date/Signature: _________________________________

Table of Contents

Other manuals for Mindray ME7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME7 and is the answer not in the manual?

Mindray ME7 Specifications

General IconGeneral
BrandMindray
ModelME7
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals