Part III Appendix
Electrical Safety Maintenance 271
Earth leakage
current
Single fault condition
(SFC)
____μA
NC: 500 μA
SFC: 1000 μA
6*
Patient leakage
current
Normal condition (NC)
BF____μA
Limit
BF application type
NC: 100 μA
SFC: 500 μA
Single fault condition
(SFC)
BF____μA
7* Leakage current on the major applied part
BF____μA
Limit
BF application type
5000 μA
8*
Patient auxiliary
current
Normal condition (NC)
BF____μA
Limit
NC:10 μA
SFC: 50 μA
Single fault condition
(SFC)
BF____μA
Note: The 4-protection ground impedance test item is applicable only to the MT3 trolley (with power) and not to
the standalone MX portable device or the MT2 trolley (without power).
Name: _____________________________Date/Signature: _________________________________