EasyManuals Logo

Mindray ME8 Service Manual

Mindray ME8
292 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Part I Service Scenarios
Installation 30
"Normal" parsing mode
The barcode is comprised of the patient ID, other IDs, patient name, DOB, and other
information.
Setup items are described as follows:
Item Description
Barcode example
Enter a barcode and the system automatically separates the barcode into the list in the
lower part based on separators, which set the start position and end position of different
information items.
Parameters After an example barcode is entered, you can modify the patient ID, other Ids,
patient name, DOB (including the year, month, and day), age, gender, and other
fields.
"Ignored item" is used to add a line of ignored item below the selected parameter
to hide unimportant information fields in the barcode.
Set the start and end positions of each item by using separators. After entering a
barcode example, select the barcode field separators. The separator drop-down list
displays only the separators that have been entered in the barcode example.
Delete: To delete only information in the "Content" column of the selected item.
Restore default: Restore the default values of parameters.
Age unit: Select the corresponding age unit, which can be year, month, or day.
Enter the custom gender symbol behind the male and female dialog boxes, for
example, male (M) and female (F).
Note:
Define the units of the DOB (year, month, and day) on the right. If the DOB is pure
digits, the system will automatically calculates the patient's age based on the DOB
provided by the patient.
Age unit Select the corresponding age unit.
Male/Female Enter the custom gender symbol.
Default item If no default item is set, the scanned information is directly input after 2D and 1D match
failures. If a default item is set, the system considers scanned information as the default
item. For example, if the default item is set to patient ID and 2D and 1D match fail after
barcode scanning, the scanned data is automatically displayed in the "Patient ID" editing
box on the patient information page.
Move Up/Down Move up or down the selected item.
Add/Delete Add or delete the selected item. Only items that are set before delivery can be added or
deleted.
Restore default Restore the default values of items.
Worklist query If a worklist server is selected, the worklist server is queried based on scanned
information.
If no worklist server is selected, a check is created on the patient information page
based on the scanned information.

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals