Description of Self Test Items C-9
Test Content
The dialog box appears on the primary monitor when performing the test. It has to change the
circle on the touch screen into green manually.
The program passes the test automatically if all circles turn into green. Click “Touch screen does
not work well” when abnormality happens in the test.
Analysis to Test Failure
N/A
Suggestion to Test Failure
If the touch screen does not work at all, reinstall the system for the abnormal drive of the touch
screen.
If some areas do not work, it is recommended to change the touch screen.
C.1.20 Z0701 Control Panel USB Interconnection Test
To p test item
N/A