Description of Self Test Items C-15
Analysis to Test Failure
The drive goes wrong if the test result appears Error.
The interconnection of TEE temperature signal goes wrong if the test appears FAIL.
Suggestion to Test Failure
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device if Error re-appears.
It is recommended to replace 4D&TEE board if the test result is FAIL.
C.1.33 Z1104 4D&TEE Board Power Supply Voltage Test
To p t e st item
DSP FPGA and 4D&TEE Board Interconnection Test
Test Content
Enable 4D drive circuits+28 V power supply to obtain the voltage value through the monitoring
circuit on the 4D board. If the voltage values is within the specifications, it indicates the voltage is
normal.
Analysis to Test Failure
The drive goes wrong if the test result appears Error.
The error may occur to 4D board + 28 V voltage if the test result is FAIL.
Suggestion to Test Failure
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device if Error re-appears.
It is recommended to replace 4D&TEE board if the test result is FAIL.
C.1.34 Z1201 DSP FPGA and ECG Board Interconnection
Test
Top test item
PC Module and DSP FPGA Interconnection Te s t
Test content
The program decides whether ECG is in the place, sends orders via ECG serial port, judges the
returned data, and parses Bootloader version, board ID and software version.
Analysis to test failure
The drive goes wrong if the test result appears Error.
ECG board is not in the place if the information “ECG board is not in the place appears or the
circuit is not in the place.
The communication between ECG board and main board goes wrong if the information “Failed to
read the data on ECG board”.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device if Error re-appears.
It is recommended to replace ECG board if the test result is FAIL.
C.1.35 Z1202 ECG Board Self Test
Top test item
DSP FPGA and ECG Board Interconnection Test.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Nuewa R9 Elite and is the answer not in the manual?

Mindray Nuewa R9 Elite Specifications

General IconGeneral
BrandMindray
ModelNuewa R9 Elite
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals