Overview
6 Operator's Manual
Item Quantity Description
Carryingcase 1 /
Tipcover 1 /
Transducercoversuite 1 OnlysuppliedwithP8-2Ts/P8-2TU.
(IncludingNon-sterileTEEcover,gel-lled
syringeandbite-guard)
Tips:
Thetipcoverenclosesandprotectsthedistalend/scanheadoftheendoscopefrombeing
exposedtomechanicalstrainduringtransportationandstorage.
2.3.2 Transducer Inspection
ThefollowinginspectionsshouldbeperformedontheTEEtransducerafterunpackingthe
contents.
• Performvisualandtactileinspection,see"3.1.1 VisualandTactileInspection".
• Performtipdeectioninspection,see"3.1.2 TipDeectionControl".
• Performbrakeinspection,see "3.1.3 Brake Operation".
• Performscanplanerotationinspection,see"3.1.4 ScanplaneRotationOperation"..
• Performleakagetest,see"3.3 ElectricalSafety".
ContactMindrayoryourlocalrepresentativeimmediatelytoreportanydamageor
discrepancies.
WARNING
To avoid injury to the patient, if any irregularity, substandard function or unsafe
condition is observed or suspected, the TEE transducer should not be used.
2.4 Structure
TEEtransducerisaphasedarraytransducer,mountedinasealedtipattheendofa
gastroscope.Thetransducerisintendedforimagingoftheheartthroughtheesophagus
andthestomach.Thearraycanberotated0°–180°.