Image Optimization  5-43 
Type  Parameter  Description 
Max 
Function: set Max as 3D image rendering mode, displays the 
maximum echo intensity in the observation direction. 
This is helpful for viewing bony structures. 
Min 
Function: set Min as 3D image rendering mode, displays the 
minimum echo intensity in the observation direction. 
This is helpful for viewing vessels and hollow structures. 
X ray 
Function: set X-ray as 3D image rendering mode. Displays the 
average value of all gray values in the ROI. 
This is used for imaging tissues with different structure inside or 
tissues with tumor. 
iLive 
Function: add the light rendering effect based on the general 
rendering effect. 
iLive cannot be selected as sub render mode. When it is set as 
main render mode, the sub render mode is disabled. 
STIC 
Based on the movements of the fetus, rebuilt and show the anatomical structure 
within a physical movement by using the interconnection between the time and the 
space. 
5.11.3.3  Static 3D Image Viewing 
Entering/ Exiting Image Viewing 
To enter image viewing 
The system enters image viewing when 3D image acquisition is finished 
Exit 
Press <Freeze> or <Update> to return to image acquisition preparation status and to exit image 
reviewing. 
 
3D/4D Scenario 
This function provides users multiple groups of preset 3D/4D parameters based on different application 
scenarios to quickly obtain expected image effect.   
 
Tap [3D/4D scenario] to select a desired scenario and subpreset (rendering mode) under each scenario 
to view images. The parameters of each rendering mode are preset to different values. 
 
Default scenarios and rendering modes of each probe and exam mode are different. You can customize 
the 3D/4D scenario in QSave. For details, please refer to “5.1.3 Quickly Saving Image Settings”. 
 
Activate MPR 
On the [VR] subtab, tap [A], [B], [C] or [VR] to activate sectional plane image (MPR) or 3D image (VR).