Image Optimization 5-141
Rotating CT/MR/PET/Freehand Cubic Image
Rotate the knob <M>, the knob <PW>, the knob <C> to cycle the CT/MR/PET/freehand cubic image
through X-axis, Y-axis and Z-axis. The 2-dimensional CT/MR/PET/freehand anatomical plane cycles
simultaneously.
The relations of X-axis, Y-axis and Z-axis are given below:
Rotate the knob <M> clockwise to cycle CT/MR/PET/freehand image clockwise through X-axis;
Rotate the knob <PW> clockwise to cycle CT/MR/PET/freehand image clockwise through Y-axis;
Rotate the knob <C> clockwise to cycle CT/MR/PET/freehand image clockwise through Z-axis;
Scrolling CT/MR/PET/Freehand Image
The description of scrolling the transverse planeis given below:
Rotate the <4D> key clockwise. The transverse plane moves towards the patients head. The R~L axis
and A~P axis on the transverse plane move towards the patients head.
R~L axis, A~P axis correspond with the axe of coronal plane and sagittal plane. R~L axis of coronal
plane and A~P axis of sagittal plane move towards the cephalosome.
Z
X
Y
Coronal plane
Sagittal plane
Transverse
plane
R
P
H
L
A
F

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona I9 and is the answer not in the manual?

Mindray Resona I9 Specifications

General IconGeneral
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
System TypeUltrasound System
TouchscreenYes
Imaging ModesColor Doppler, Power Doppler
Advanced ImagingV Flow
ConnectivityDICOM, USB, HDMI, Ethernet
ProbesConvex, Linear, Phased Array, Volume
PortabilityCart-based
Storage Temperature-20°C to 55°C
Humidity Range30% to 80% (non-condensing)
Display23.8-inch high-resolution LED monitor
Transducer Ports4